User:PAR/Work4
From Wikipedia, the free encyclopedia
In particle and nuclear detector systems the dead time is the time after each event, during which the system is not able to record another event if it happens. An everyday life example of this is what happens when someone takes a photo using a flash - the eyes need time to readjust, leaving you unable to see for a few seconds.
The total dead time of a detection system is usually due to the contributions of the intrinsic dead time of the detector (for example the drift time in a gaseous ionization detector), of the analog front end (for example the shaping time of a spectroscopy amplifier) and of the DAQ (the conversion time of the ADCs and the readout and storage times).
The intrinsic dead time of a detector is often due to its physical characteristics; for example a spark chamber is "dead" until the potential between the plates recovers above a high enough value. In other cases the detector, after a first event, is still "live" and does produce a signal for the successive event, but the signal is such that the detector readout is unable to discriminate and separate them, resulting in an event loss or in a so called "pile-up" event where, for example, a (possibly partial) sum of the deposited energies from the two events is recorded instead. In some cases this can be minimised by an appropriate design, but often only at the expense of other properties like energy resolution.
The analog electronics can also introduce dead time; in particular a shaping spectroscopy amplifier needs to integrate a fast rise, slow fall signal over the longest possible time (usually from .5 up to 10 microseconds) to attain the best possible resolution, such that the user needs to choose a compromise between event rate and resolution.
Trigger logic is another possible source of dead time; beyond the proper time of the signal processing, spurious triggers caused by noise need to be taken into account.
Finally, digitisation, readout and storage of the event, especially in detection systems with large number of channels like those used in modern High Energy Physics experiments, also contribute to the total dead time. To alleviate the issue, medium and large experiments use sophisiticated pipelining and multi-level trigger logic to reduce the readout rates[1].
From the total time a detection system is running, the dead time must be subtracted to obtain the live time of the experiment.
[edit] Paralizable and non-paralizable behaviour
A detector, or detection system, can be characterized by a paralizable or non-paralizable behaviour.
In a non-paralizable detector, an event happening during the dead time since the previous event is simply lost, so that with an increasing event rate the detector will reach a saturation rate equal to the inverse of the dead time.
In a paralizable detector, an event happening during the dead time since the previous one will not just be missed, but will restart the dead time, so that with increasing rate the detector will reach a saturation point where it will be incapable of recording any event at all.
A semi-paralizable detector exhibits an intermediate behaviour, in which the event arriving during dead time does extend it, but not by the full amount, resulting in a detection rate that decreases when the event rate approaches saturation.
[edit] Analysis
It will be assumed that the events are occurring randomly with an average frequency of f. That is, they constitute a Poisson process. The probability that an event will occur in an infinitesimal time interval dt is then f dt. The probability P(t) that an event will occur at time t to t+dt with no events occurring between t=0 and time t can be shown to be given by the exponential distribution (Lucke, 1974):
The expected time between events is then
[edit] Non-paralizable analysis
For the non-paralizable case, with a dead time of τ, the probability of measuring an event between t=0 and t = τ is zero. Otherwise the probabilities of measurement are the same as the event probabilities. The probability of measuring an event at time t with no intervening measurements is then given by
for 
for 
The expected time between measurements is then
In other words, if Nm counts are recorded during a particular time interval, the actual number of events (N) may be estimated by





